中华口腔医学研究杂志(电子版) ›› 2023, Vol. 17 ›› Issue (04) : 285 -291. doi: 10.3877/cma.j.issn.1674-1366.2023.04.009

论著

新型硅橡胶印模消毒流程对模型精度的影响
张佳园, 魏凌飞, 刘晶, 邬春兰, 叶丽娟, 于德栋()   
  1. 上海交通大学医学院附属第九人民医院口腔第二门诊部,上海交通大学口腔医学院,国家口腔医学中心,国家口腔疾病临床医学研究中心,上海市口腔医学重点实验室,上海市口腔医学研究所,上海 201999;杭州医学院存济口腔医学院 杭州医学院附属口腔医院,杭州 310012
    滨州医学院附属烟台口腔医院口腔种植科,烟台 264003
    上海交通大学医学院附属第九人民医院口腔第二门诊部,上海交通大学口腔医学院,国家口腔医学中心,国家口腔疾病临床医学研究中心,上海市口腔医学重点实验室,上海市口腔医学研究所,上海 201999
  • 收稿日期:2022-11-09 出版日期:2023-08-01
  • 通信作者: 于德栋

Effect of new silicone rubber impression disinfection process on model accuracy

Jiayuan Zhang, Lingfei Wei, Jing Liu, Chunlan Wu, Lijuan Ye, Dedong Yu()   

  1. Department of Second Dental Center, Shanghai Ninth People′s Hospital, Shanghai Jiao Tong University School of Medicine; College of Stomatology, Shanghai Jiao Tong University; National Center for Stomatology; National Clinical Research Center for Oral Diseases; Shanghai Key Laboratory of Stomatology, Shanghai Research Institute of Stomatology, Shanghai 201999, China; Hangzhou Medical College·Cunji College of Stomatology, Affiliated Stomatological Hospital of Hangzhou Medical College, Hangzhou 310012, China;
    Yantai Stomatological Hospital Affiliated to Binzhou Medical College, Department of Oral Implantation, Yantai 264003, China
    Department of Second Dental Center, Shanghai Ninth People′s Hospital, Shanghai Jiao Tong University School of Medicine; College of Stomatology, Shanghai Jiao Tong University; National Center for Stomatology; National Clinical Research Center for Oral Diseases; Shanghai Key Laboratory of Stomatology, Shanghai Research Institute of Stomatology, Shanghai 201999, China;
  • Received:2022-11-09 Published:2023-08-01
  • Corresponding author: Dedong Yu
  • Supported by:
    National Natural Science Foundation of China(52175422, 32101094); Shanghai Municipal Health Commission Health Industry Clinical Research Project(202240194); "Cross" Research Fund Project of the Ninth People′s Hospital Affiliated to the Medical College of Shanghai Jiaotong University(JYJC202114); 2021 "Seed Fund" Project(JYZZ145)
引用本文:

张佳园, 魏凌飞, 刘晶, 邬春兰, 叶丽娟, 于德栋. 新型硅橡胶印模消毒流程对模型精度的影响[J]. 中华口腔医学研究杂志(电子版), 2023, 17(04): 285-291.

Jiayuan Zhang, Lingfei Wei, Jing Liu, Chunlan Wu, Lijuan Ye, Dedong Yu. Effect of new silicone rubber impression disinfection process on model accuracy[J]. Chinese Journal of Stomatological Research(Electronic Edition), 2023, 17(04): 285-291.

目的

评价新型口腔印模清洗消毒流程对于硅橡胶材料牙科印模精度的影响。

方法

对16名青年志愿者制取硅橡胶口腔印模(每人上下颌各2次),采用口内扫描仪对牙列进行扫描(每人上下颌各1次),将得到的扫描数据作为参考数据。每位志愿者的2对硅橡胶印模分别编号为1组和2组,其中1组不消毒,2组按照新型硅橡胶印模清洗消毒流程对口腔印模进行处理。将受试者的所有口腔印模进行标准的石膏灌注和脱模,并通过仓式扫描仪对石膏模型进行扫描,获得测试数据。分别配准对比分析每名志愿者的测试数据与参考数据,将误差的均方根(RMS)作为评价测试模型与参考模型间偏差的指标参数。比较各组模型之间的数值偏差并对三维偏差色谱图进行可视化分析,从多个维度分析比较模型精度差异。

结果

上颌牙列消毒前后模型RMS偏差分别为(0.41 ± 0.21)和(0.49 ± 0.19) mm;下颌牙列消毒前后模型RMS偏差分别为(0.46 ± 0.26)和(0.38 ± 0.13) mm。硅橡胶印模上下颌牙列消毒前后模型精度差异均无统计学意义(t上颌 = -1.139,P上颌 = 0.272;t下颌 = 1.113,P下颌 = 0.283)。

结论

该新型硅橡胶口腔印模清洗消毒流程对模型精度没有显著影响,在临床实践操作中具有良好的可行性。

Objective

To evaluate the effect of a new oral impression cleaning-disinfection process on the accuracy of dental impressions made of silicone rubber materials.

Methods

Silicone rubber oral impressions were made on 16 young volunteers (upper and lower jaws for each person twice), and the dentition was scanned using an intraoral scanner (upper and lower jaws for each person twice). The obtained scanning data were used as reference data. The two pairs of silicone rubber impressions of each volunteer were numbered as Group 1 and Group 2. The impressions in Group 1 were not disinfected, and those in Group 2 were treated by the new silicone rubber impression cleaning and disinfection process. All the oral impressions of the subjects were perfused with plaster and demolded in the standard way, and the plaster models were scanned using a laboratory scanner to obtain test data. Registration and comparative analysis were carried out for the test and reference data of each volunteer separately, and the root mean square (RMS) of the error was taken as an indicator parameter to evaluate the deviation between the test and reference models. The numerical deviation between the two groups was compared. Visual analysis on the three-dimensional deviation chromatogram was performed, and the difference in model accuracy from multiple dimensions was analyzed.

Results

The RMS deviations of the models before and after disinfection of the maxillary dentition were (0.41 ± 0.21) mm and (0.49 ± 0.19) mm. The RMS deviations of the models before and after disinfection of the mandibular dentition were (0.46 ± 0.26) mm and (0.38 ± 0.13) mm. There was no significant difference in the accuracy of the model of the upper and lower dentitions made from the silicone rubber impression before and after disinfection (tupper jaw = -1.139, Pupper jaw = 0.272; tlower jaw = 1.113, Plower jaw = 0.283) .

Conclusions

The new silicone rubber oral impression cleaning-disinfection process had no significant impact on the accuracy of model. This new disinfection process had good feasibility in clinical practice.

图1 数字化模型的数据处理(以上颌为例) A:将数字化模型沿龈缘线分界进行剪切,切除临床牙冠以外部分;B:剪切后得到的全牙弓数字化模型。
图2 参考模型与测试模型利用右侧上颌第一、二磨牙解剖标志点进行对齐(以上颌为例)
表1 硅橡胶印模材料上下颌牙列消毒前后模型间RMS偏差
图3 硅橡胶材料消毒前后牙列偏差三维分析色谱图 A:上颌未消毒;B:上颌消毒;C:下颌未消毒;D:下颌消毒。
[1]
赵铱民. 口腔修复学[M]. 8版. 北京:人民卫生出版社,2020:34-35.
[2]
赵信义. 口腔材料学[M]. 6版. 北京:人民卫生出版社,2020:105-116.
[3]
Hardan L, Bourgi R, Cuevas-Suárez CE,et al. Disinfection procedures and their effect on the microorganism colonization of dental impression materials:A systematic review and meta-analysis of in vitro studies[J]. Bioengineering(Basel)2022,9(3):123. DOI:10.3390/bioengineering9030123.
[4]
Singla Y, Pachar RB, Poriya S,et al. Evaluation of the efficacy of different mixing techniques and disinfection on microbial colonization of polyether impression materials:A comparative study[J]. J Contemp Dent Pract2018,19(3):296-300. DOI:10.5005/jp-journals-10024-2256.
[5]
Gumru B, Tarcin B, Idman E. Cross-contamination and infection control in intraoral digital imaging:A comprehensive review[J]. Oral Radiol2021,37(2):180-188. DOI:10.1007/s11282-020-00452-z.
[6]
Varvara G, Murmura G, Sinjari B,et al. Evaluation of defects in surface detail for monophase,2-phase,and 3-phase impression techniques:An in vitro study[J]. J Prosthet Dent2015,113(2):108-113. DOI:10.1016/j.prosdent.2014.08.007.
[7]
Kampf G, Todt D, Pfaender S,et al. Persistence of coronaviruses on inanimate surfaces and their inactivation with biocidal agents[J]. J Hosp Infect2020,104(3):246-251. DOI:10.1016/j.jhin.2020.01.022.
[8]
Cervino G, Fiorillo L, Herford AS,et al. Alginate materials and dental impression technique:A current state of the art and application to dental practice[J]. Mar Drugs2018,17(1):18. DOI:10.3390/md17010018.
[9]
Flügge T, van der Meer WJ, Gonzalez BG,et al. The accuracy of different dental impression techniques for implant-supported dental prostheses:A systematic review and meta-analysis[J]. Clin Oral Implants Res2018,29(Suppl 16):374-392. DOI:10.1111/clr.13273.
[10]
Mathew S, Alani MM, Nair KN,et al. Radiofrequency glow discharge as a mode of disinfection for elastomeric impression materials[J]. J Contemp Dent Pract2017,18(2):131-136. DOI:10.5005/jp-journals-10024-2003.
[11]
Jayaraman S, Singh BP, Ramanathan B,et al. Final-impression techniques and materials for making complete and removable partial dentures[J]. Cochrane Database Syst Rev2018,4(4):CD012256. DOI:10.1002/14651858.CD012256.pub2.
[12]
Choudhury GK, Chitumalla R, Manual L,et al. Disinfectant efficacy of 0.525% sodium hypochlorite and epimax on alginate impression material[J]. J Contemp Dent Pract2018,19(1):113-116. DOI:10.5005/jp-journals-10024-2222.
[13]
WHO. Cleaning and disinfection of environmental surfaces in the context of COVID-19[EB/OL].(2020-05-16)[2021-01-22].

URL    
[14]
刘永健,李天一,邓永强,等. 次氯酸钠消毒剂灭活新冠病毒的实验研究[J]. 军事医学2022,46(3):192-196. DOI:10.7644/j.issn.1674-9960.2022.03.007.
[15]
Schaefer O, Watts DC, Sigusch BW,et al. Marginal and internal fit of pressed lithium disilicate partial crowns in vitro:A three-dimensional analysis of accuracy and reproducibility[J]. Dent Mater2012,28(3):320-326. DOI:10.1016/j.dental.2011.12.008.
[16]
李雅,陈志宇,张辉伟. 不同扫描方法重建上颌牙列数字化模型的三维对比分析[J]. 现代口腔医学杂志2015,29(3):157-160. DOI:CNKI:SUN:XDKY.0.2015-03-011.
[17]
柯雯,屈直. 不同印模材料取模的三维精度比较[J]. 暨南大学学报(自然科学与医学版)2017,38(3):269-273. DOI:10.11778/j.jdxb.2017.03.014.
[18]
张翠云. 不同种植印模方式和印模材料精度的体外研究[D]. 福州:福建医科大学,2018.
[19]
周甜,解保生,蒋娟娜. 三种消毒剂浸泡消毒对硅橡胶印模尺寸稳定性影响的研究[J]. 实用口腔医学杂志2020,36(2):295-297. DOI:10.3969/j.issn.1001-3733.2020.02.037.
[20]
徐迪,魏冬豪,张亚池,等. 以苄索氯胺和异丙醇为主要有效成分的消毒剂对牙科印模精度的影响[J]. 北京大学学报(医学版)2020,52(6):1112-1116. DOI:10.19723/j.issn.1671-167X.2020.06.021.
[21]
武明彤,唐素霞,彭玲燕,等. 仿真头颅模型口内和手持牙颌模型条件下四种口内扫描仪全牙弓扫描时间和精度的对比研究[J]. 中华口腔医学杂志2021,56(6):570-575. DOI:10.3760/cma.j.cn112144-20210221-00082.
[22]
Cavalli A, Marinaro M, Desario C,et al. In vitro virucidal activity of sodium hypochlorite against canine parvovirus type 2[J]. Epidemiol Infect2018,146(15):2010-2013. DOI:10.1017/S0950268818002431.
[23]
Rutala WA, Weber DJ. Uses of inorganic hypochlorite (bleach) in health-care facilities[J]. Clin Microbiol Rev1997,10(4):597-610. DOI:10.1128/CMR.10.4.597.
[24]
AlZain S. Effect of 0.5% glutaraldehyde disinfection on surface wettability of elastomeric impression materials[J]. Saudi Dent J2019,31(1):122-128. DOI:10.1016/j.sdentj.2018.10.002.
[25]
Hatanaka N, Xu B, Yasugi M,et al. Chlorine dioxide is a more potent antiviral agent against SARS-CoV-2 than sodium hypochlorite[J]. J Hosp Infect2021,118:20-26. DOI:10.1016/j.jhin.2021.09.006.
[26]
Balkenhol M, Ferger P, Wöstmann B. Dimensional accuracy of 2-stage putty-wash impressions:Influence of impression trays and viscosity[J]. Int J Prosthodont2007,20(6):573-575. DOI:10.1111/j.1365-263X.2007.00870.x
[27]
Kerr M, Park N, Leeson D,et al. Dimensional accuracy of microcomputed tomography-scanned half-arch impressions[J]. J Prosthet Dent2019,121(5):797-802. DOI:10.1016/j.prosdent.2018.08.010.
[28]
颉慧菲. 储存环境及时间对硅橡胶印模精度影响的实验研究[D]. 西安:第四军医大学,2010. DOI:10.7666/d.d219628.
[29]
Martins F, Branco P, Reis J,et al. Dimensional stability of two impression materials after a 6-month storage period[J]. Acta Biomater Odontol Scand2017,3(1):84-91. DOI:10.1080/23337931.2017.1401933.
[30]
Hiraguchi H, Kaketani M, Hirose H,et al. Surface porosity of stone casts resulting from immersion of addition silicone rubber impressions in disinfectant solutions[J]. Int J Prosthodont2014,27(6):567-569. DOI:10.11607/ijp.3967.
No related articles found!
阅读次数
全文


摘要